Drop Test of Led Mining Cap Lamp
In order to check the quality of the mining cap lamp, Brando did a drop test of the explosion-proof mining cap lamp. this is our test report:
Drop Test | ||||
sample status /amount | 3 samples in working condition | |||
experimental objective | Expose the shell material damage and failure caused by falling of the prototype | |||
test method | 1. After setting to the state of turning on the light, the drop times of each face shall be 1 for the corresponding height of the 6 faces of back, front, left side, right side, bottom, top and 4 angles; 2.Each face falls backward must carry on the appearance, the structure, the function inspection to the entire lamp; 3.The falling ground is 100mm concrete slab. | |||
judgment basis | Product appearance, structure and function are normal 1.Function check: whether the charging is normal, confirm whether the charging, screen, button switch can be used normally; 2.Structural inspection: structural parts, logo, etc. No ungumming and other inconsistencies with pre-test conditions; 3.Appearance inspection: no cracks on the surface of the shell and other phenomena inconsistent with the state before testing; 4.Sample memory, clock no abnormal (memory no loss; Clock without chaos, reset, lead, lag, etc.). | |||
Height(1m) | ||||
surface | appearance | function | structure | remark |
reverse side | OK | OK | OK | |
front | OK | OK | OK | |
left surface | OK | OK | OK | |
The right side | OK | OK | OK | |
underside | OK | OK | OK | |
top surface | OK | OK | OK | |
left bottom | OK | OK | OK | |
bottom right corner | OK | OK | OK | |
top left corner | OK | OK | OK | |
top right corner | OK | OK | OK |